Your browser doesn't support javascript.
loading
Fatigue property analysis of prosthesis of hip joint with two different materials / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 73-76, 2015.
Artículo en Chino | WPRIM | ID: wpr-266724
ABSTRACT
Total hip replacement (THR) is replacing the prosthesis stem similar to human bone that takes advantage of the material with both good mechanical properties and biocompatibility to the damaged articular surface. Thus it can not only alleviate or even eliminate the pain but also effectively maintain the joint stability and freedom and restore its normal performance. Finite element analysis was used in this study to establish a 3D model of artificial hip stem, and explore its fatigue properties of different materials to ensure the safety and reliability. The calculating obtained two results of different metal hip prosthesis, including lifetime and deformation. The minimum service life of titanium prosthesis reaches 568 million times, which satisfies ISO standards, while the stainless steel does not suit to be a prosthesis material.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Acero Inoxidable / Titanio / Ensayo de Materiales / Falla de Prótesis / Reproducibilidad de los Resultados / Artroplastia de Reemplazo de Cadera / Análisis de Elementos Finitos / Articulación de la Cadera / Prótesis de Cadera / Modelos Anatómicos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2015 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Acero Inoxidable / Titanio / Ensayo de Materiales / Falla de Prótesis / Reproducibilidad de los Resultados / Artroplastia de Reemplazo de Cadera / Análisis de Elementos Finitos / Articulación de la Cadera / Prótesis de Cadera / Modelos Anatómicos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2015 Tipo del documento: Artículo