Your browser doesn't support javascript.
loading
Fast gamma index calculation method in dose distribution comparison / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 550-554, 2012.
Artículo en Chino | WPRIM | ID: wpr-271735
ABSTRACT
As a method of dosimetric verification in radiotherapy, gamma index has been widely used for evaluating dose distribution in research and clinical cases. However, for three-dimensional dose distributions, gamma index calculation is very time consuming for the computers. In this paper, based on a pre-sorting technique, we implement a parallel computing algorithm of gamma index on graphic processing unit (GPU). Dose comparisons are performed for seven cases to test our new implementation. It was shown that the GPU-based gamma index calculations achieved a speedup of ten-folds in comparison with corresponding CPU implementation without losing accuracy. The result showed that utilizing GPU parallel computing to speed up gamma index calculations could be reliable and efficient in the implementation.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Radiometría / Dosificación Radioterapéutica / Algoritmos / Gráficos por Computador / Planificación de la Radioterapia Asistida por Computador / Programas Informáticos / Métodos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2012 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Radiometría / Dosificación Radioterapéutica / Algoritmos / Gráficos por Computador / Planificación de la Radioterapia Asistida por Computador / Programas Informáticos / Métodos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2012 Tipo del documento: Artículo