Simultaneous determination of 3-chlorotyrosine and 3-nitrotyrosine in human plasma by direct analysis in real time-tandem mass spectrometry
Acta Pharmaceutica Sinica B
;
(6): 482-486, 2015.
Artículo
en Inglés
| WPRIM
| ID: wpr-310003
ABSTRACT
A novel method for the simultaneous determination of 3-nitrotyrosine (NT) and 3-chlorotyrosine (CT) in human plasma has been developed based on direct analysis in real time-tandem mass spectrometry (DART-MS/MS). Analysis was performed in the positive ionization mode using multiple reaction monitoring (MRM) of the ion transitions at m/z 216.2/170.1 for CT, m/z 227.2/181.1 for NT and m/z 230.2/184.2 for the internal standard, d (3)-NT. The assay was linear in the ranges 0.5-100 μg/mL for CT and 4-100 μg/mL for NT with corresponding limits of detection of 0.2 and 2 μg/mL. Intra- and inter-day precisions and accuracies were respectively <15% and ±15%. Matrix effects were also evaluated. The method is potentially useful for high throughput analysis although sensitivity needs to be improved before it can be applied in clinical research.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Inglés
Revista:
Acta Pharmaceutica Sinica B
Año:
2015
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS