Investigation of Elekta linac characteristics for VMAT / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 22-24, 2012.
Artículo
en Chino
| WPRIM
| ID: wpr-325944
ABSTRACT
<p><b>UNLABELLED</b>The aim of this study is to investigate the characteristics of Elekta delivery system for volumetric modulated arc therapy (VMAT). Five VMAT plans were delivered in service mode and dose rates, and speed of gantry and MLC leaves were analyzed by log files. Results showed that dose rates varied between 6 dose rates. Gantry and MLC leaf speed dynamically varied during delivery.</p><p><b>CONCLUSION</b>The technique of VMAT requires linac to dynamically control more parameters, and these key dynamic variables during VMAT delivery can be checked by log files. Quality assurance procedure should be carried out for VMAT related parameter.</p>