Your browser doesn't support javascript.
loading
Comparative test on puncture coring of two different needles used with the implantable drug-supplying device / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 445-447, 2010.
Artículo en Chino | WPRIM | ID: wpr-330539
ABSTRACT
A comparison test on the number of puncture coring is conducted using the normal needle and the Huber needle in the injection area of the implantable drug-supplying device separately. The result indicates that the number of coring using the Huber needle is much less than that using the normal needle during the puncturing. So it is suggested to popularize the Huber needle in the drug transfusion of the implantable drug-supplying device.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Bombas de Infusión Implantables / Clasificación / Agujas Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2010 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS