Determination of1-bromopropane in the workplace air by GC-FID / 中华劳动卫生职业病杂志
Chinese Journal of Industrial Hygiene and Occupational Diseases
;
(12): 467-469, 2013.
Artículo
en Chino
| WPRIM
| ID: wpr-343649
ABSTRACT
<p><b>OBJECTIVE</b>To establish a method for determination 1-bromopropane (1-BP) in workplace air by gas chromatography-flame ionization detector (GC-FID).</p><p><b>METHOD</b>1-bromopropane in workplace air was collected with activated charcoal tube, then desorbed by carbon disulfide and determined by GC-FID. 1-bromopropane was quantitatively measured using retention time and peak area.</p><p><b>RESULTS</b>Linear regression formula was Y = 3353.4x-10064 in a range of 2.50 ∼ 500.00 µg/ml with regression coefficient R = 0.9998. Detection limit was 0.25 µg/ml and the lowest detection concentration of 1-brmopropane in air was 0.14 mg/m(3) (at air volume 1.8L). The mean recoveries of 1-BP were between 96.8% and 102.6%, and relative standard deviation of inter and intra-assay was less than 10%. The average desorption efficiencies were between 93.2% and 104.4%. The samples in activated charcoal tube could be stably stored for 5 days at room temperature.</p><p><b>CONCLUSION</b>The method could be feasible to determine 1-bromopropane in workplace air.</p>
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Cromatografía de Gases
/
Lugar de Trabajo
/
Aire
/
Contaminantes Ocupacionales del Aire
/
Hidrocarburos Bromados
/
Métodos
Idioma:
Chino
Revista:
Chinese Journal of Industrial Hygiene and Occupational Diseases
Año:
2013
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS