Your browser doesn't support javascript.
loading
A real-time flow velocity estimation applied to ultrasound color display / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 530-533, 2007.
Artículo en Chino | WPRIM | ID: wpr-357658
ABSTRACT
In this paper the theory of autocorrelation algorithm for color flow mapping is analyzed and a new way for ultrasound color flow velocity estimation and real-time display is proposed. The method sets up a mapping table which directly relates to the dynamic display range and has only 256 entries using an inverse mapping method instead of calculating the arctangent value directly. This method is ideal for software implementation and offers an interactive way to the user for changing the dynamic range of flow velocity and thus to increasing the display resolution.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fisiología / Reología / Velocidad del Flujo Sanguíneo / Algoritmos / Flujo Pulsátil / Ultrasonografía Doppler en Color / Métodos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2007 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fisiología / Reología / Velocidad del Flujo Sanguíneo / Algoritmos / Flujo Pulsátil / Ultrasonografía Doppler en Color / Métodos Límite: Humanos Idioma: Chino Revista: Journal of Biomedical Engineering Año: 2007 Tipo del documento: Artículo