Your browser doesn't support javascript.
loading
Mechanical Valve Stuck in the Mitral Position in a Patient with Antiphospholipid Syndrome / 日本心臓血管外科学会雑誌
Japanese Journal of Cardiovascular Surgery ; : 414-417, 2000.
Artículo en Japonés | WPRIM | ID: wpr-366628
ABSTRACT
A 69-year-old woman, who had undergone mitral valve replacement, developed acute congestive heart failure and was transferred to our institution. Cineradiography demonstrated that two leaflets of the St. Jude Medical valve were stuck in a closed position. Emergency redo mitral valve replacement was performed with a CarboMedics valve. Postoperative hematological studies yielded a diagnosis of antiphospholipid syndrome. Although postoperative anticoagulant therapy was performed more carefully than usual, the prosthesis became stuck again. Therefore, a third operation was performed using a tissue prosthesis. We concluded that mitral valve plasty should be a first option for patients with antiphospholipid syndrome undergoing mitral valve surgery. Should prosthetic valve replacement be required, a tissue prosthesis would be best.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Japonés Revista: Japanese Journal of Cardiovascular Surgery Año: 2000 Tipo del documento: Artículo

Similares