Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens
Annals of Laboratory Medicine
;
: 341-347, 2015.
Artículo
en Inglés
| WPRIM
| ID: wpr-36804
ABSTRACT
BACKGROUND:
The identification of in vitro hemolysis (IVH) using a hematology analyzer is challenging because centrifugation of the specimens cannot be performed for cell counts. In the present study, we aimed to develop a scoring system to help identify the presence of hemolysis in anticoagulated blood specimens.METHODS:
Thirty-seven potassium EDTA anticoagulated blood specimens were obtained, and each specimen was divided into 3 aliquots (A, B, and C). Aliquots B and C were mechanically hemolyzed by aspirating 2 and 5 times, respectively, using a 27-gauge needle and then tested; aliquot A was analyzed immediately without any hemolysis. After the cells were counted, aliquots B and C were centrifuged and the supernatants were tested for the hemolytic index and lactate dehydrogenase levels.RESULTS:
The 4 hematologic parameters were selected and scored from 0 to 3 as follows or =38.5 for mean cell hemoglobin concentration (MCHC, g/dL); or =0.04 for red blood cell ghosts (10(12)/L); or =1.31 for difference value (g/dL) of measured hemoglobin and calculated hemoglobin; and or =3.35 for difference value (g/dL) of MCHC and cell hemoglobin concentration mean. The hemolysis score was calculated by adding all the scores from the 4 parameters. At the cutoff hemolysis score of 3, the IVH of aliquots B and C were detected as 64.9% and 91.9%, respectively.CONCLUSIONS:
The scoring system might provide effective screening for detecting spurious IVH.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Hemoglobinas
/
Recolección de Muestras de Sangre
/
Ácido Edético
/
Hemólisis
/
Anticoagulantes
Tipo de estudio:
Estudio pronóstico
Límite:
Humanos
Idioma:
Inglés
Revista:
Annals of Laboratory Medicine
Año:
2015
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS