Quality control testing and evaluation for digital mammography / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 89-92, 2010.
Artículo en Chino | WPRIM | ID: wpr-390721
ABSTRACT
Objective To carry out the quality control testing and evaluation for three digital mammography systems.Methods The performance of three digital mammography systems was assessed by applying methods recommended in the European guidelines for quality assurance in breast cancer screening and diagnosis and Chinese specification for testing of quality control in X-ray mammography.The performance of X-ray generator of three digital mammography systems were tested and evaluated.CDMAM 3.4 phantom with four different thickness(30,40,50,60 mm) were exposured in DR,PCM,and CR system,respectively.The average glandular dose (AGD) value was measured and image quality figure (IQF) analysis was performed in each thickness.Results The X-ray machine performance of DR and CR was in accordance with existing standard,however the standard was inappropriate to evaluate part of X-ray machine performance of PCM system.The AGDs for system DR were 1.20,1.42,1.75 and 2.20 mGy for 30,40,50 and 60 mm PMMA thickness,respectively.The respective AGDs for system PCM and CR were 0.82,1.19,1.33,1.70 mGy and 0.59,0.88,1.47,2.19 mGy.For the same phantom thickness sequence,the IQFs were 21.36,21.57,27.25 and 30.58 for system DR,28.02,29.10,35.90,and 41.24 for system PCM,whereas they were 39.78,39.30,43.85 and 48.08 for system CR.Conclusions The AGDs of all three systems were in accordance with the values recommended in European guideline.The AGD and IQF could provide an effective way for performance assessment and constancy checks for digital mammography systems.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2010 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2010 Tipo del documento: Artículo