Dual energy CT intracranial angiography: image quality, radiation dose and initial application results / 中华放射学杂志
Chinese Journal of Radiology ; (12): 725-729, 2009.
Artículo en Chino | WPRIM | ID: wpr-394143
ABSTRACT
Objective To assess the clinical value of dual-energy intracranial CT angiography (CTA).Methods Forty-one patients suspected of intracranial vascular diseases underwent dual-energy intracranial CT angiography, and 41 patients who underwent conventional subtraction CT were enrolled as the control group.Image quality of intracranial and skull base vessels and radiation dose between dual-energy CTA and conventional subtraction CTA were compared using two independent sample nonparametrie test and independent-samples t test, respectively.Prevalence and size of lesions detected by dual-energy CTA and digital subtraction CTA were compared using paired-samples t test and Spearman correlative analysis. Results The percentage of image quality scored 5 was 70.7% (29/41) for dual-energy CTA and 75.6% (31/41) for conventional subtraction CTA.There was no significant difference between the two groups(Z= -0.455, P=0.650).Image quality of vessels at the skull base in conventional subtraction CTA was superior to that in dual-energy CTA, especially for the petrosal and syphon segment (Z=-4.087, P= 0.000).Radiation exposure of dual energy CTA and conventional CTA were (396.54±17.43) and (1090.95±114.29) mGy · cm respectively.Radiation exposure was decreased by 64% (t=-38.52, P=0.000) by dual energy CTA compared with conventional subtraction CTA.Out of the 41 patients,19 patients were diagnosed as intracranial aneurysm, 2 patients as arteriovenous malformation (AVM), 3 patients with Moya-moya's disease, and the remaining 17 patients with negative results.Nine patients with intracranial aneurysm, 2 patients with AVM, 3 patients with Moya-moya's disease, and 2 patients with negative findings underwent DSA or operation, with concordant findings from both techniques.Diameter of aneurysm neck, long axis and minor axis by dual-energy CTA was (2.90±1.61), (5.23±1.68) and (3.83±1.69) nun, respectively; Diameter of aneurysm neck, long axis and minor axis by DSA was (2.95±1.71), (5.10±1.60) ,(3.83±1.65) nan,respectively.There was no significant difference for the diameters of aneurysm between dual energy CTA and DSA ((t=-0.734,1.936,0.12.5 respectively, P=0.482,0.085,0.903 respectively), and good correlation was found between diameter measurements using the two techniques(r=0.964,0.976,0.973, respectively, all P=0.000) Conclusions Compared with conventional subtraction CTA, dual energy CTA has good image quality for intracranial vessels; however, image quality of the skull base vessels is worse, especially for the petrosal and syphon segment.Dual energy CTA has decreased radiation dose and a high diagnostic accuracy, being a practical imaging madality for diagnosis of intracranial vascular lesions.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiology Año: 2009 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiology Año: 2009 Tipo del documento: Artículo