Clinical analysis of inappropriate shock of implantable cardioverter defibrillators / 中华内科杂志
Article en Zh | WPRIM | ID: wpr-397130
Biblioteca responsable: WPRO
ABSTRACT
Objective Implantable cardioverter defibrillator(ICD)can effectively treat lifethreatening ventricular arrhythmias.The most common side effect is inappropriate discharge.This study analyzes the incidence and causes of inappropriate discharges of ICD in our hospital.Methods Forty.threepatients implanted with ICD in our hospital from November 2001 to October 2007 were involved in our study.Patients were followed-up regularly.All episodes recorded and stored in the ICD were analyzed.Results Seven of the 43 patients underwent ninety-six inappropriate discharges.Inappropriate discharges in six patients were caused by supraventricular tachyarrhythmias(SVT).In one patient the discharge was caused by noise.Most inappropriate discharges occurred in the first year after implantation.The history of atrial fibrillation before implantation is an independent predictor of inappropriate discharges.Conclusions The incidence of inappropriate discharge is 16.3%in our study and the most common cause is SVT.Most inappropriate discharges occur in the first year after implantation.Patients with atrial fibrillation history have a higher risk of inappropriate discharges.
Palabras clave
Texto completo: 1 Índice: WPRIM Tipo de estudio: Prognostic_studies Idioma: Zh Revista: Chinese Journal of Internal Medicine Año: 2009 Tipo del documento: Article