Evaluate the applicability of the international criteria for blood smear review on the hematology analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 432-433, 2010.
Artículo en Chino | WPRIM | ID: wpr-402685
ABSTRACT
Objective To evaluate the applicability of the international criteria for blood smear review on the SysmexXE-2100 hematology analyzer.Methods 300 blood samples were tested by hematology analyzer Sysmex-XE2100,then it were divided into two groups,that is,sieving group and sieved group according to the criteria of international boold smear review,then blood smears were reviewed with optical microscope.Results Among 300 samples,,54 cases were true positive(18%),true negative was 64%,48 cases were false positive(16%),four 6cases were false-negative (2%),the accuracy rate was 82%.Conclusion Performance of microscopic observation after analyzer detection according to the international criteria for blood smear review presented good efficiency,it will show better practical value analysising with the clinical condition.