Continuous Detection Method for Ractopamine with Surface Plasmon Resonance Biosensor / 分析化学
Article en Zh | WPRIM | ID: wpr-403203
Biblioteca responsable: WPRO
ABSTRACT
The interaction between the anti-Ractopamine (Rac) monoclonal antibody and the Rac derivation immobilized on the sensor chip surface was studied with surface plasmon resonance (SPR) biosensor. A continuous detection method based on the linear response during the association phase was developed. The sensor chip surface was regenerated after several tests performed continuously, thus the detection step was simplified and the life span of the chip was prospected to be prolonged. The detection was performed as an inhibitive immunoassay. The mixture of anti-Rac monoclonal antibody and the sample flowed over the surface with Rac derivation was immobilized. The relative response was in inversely proportion to the concentration of Rac. The detection limit was less than 4 μg/L with a detection time of 15 min.
Palabras clave
Texto completo: 1 Índice: WPRIM Tipo de estudio: Diagnostic_studies Idioma: Zh Revista: Chinese Journal of Analytical Chemistry Año: 2010 Tipo del documento: Article