Comparison of multi-slice CT coronary artery imaging with coronary angiography / 中国组织工程研究
Chinese Journal of Tissue Engineering Research
;
(53): 8792-8796, 2008.
Artículo
en Chino
| WPRIM
| ID: wpr-406819
ABSTRACT
BACKGROUND:
Primary studies suggest that coronary artery stenosis is highly exactly shown by 16-slice spiral CT coronary artery imaging.OBJECTIVE:
To compare the accuracy and limitation between coronary angiography and multi-slice computed tomography (MSCT) coronary artery imaging to diagnose moderate and severe coronary artery stenosis. DESIGN, TIME ANDSETTING:
Clinical diagnostic study based on gold standard, which was carried out in the Department of Cardiology, Xuanwu Hospital, Capital Medical University from June 2005 to March 2006.PARTICIPANTS:
Twenty-eight patients with suspected coronary arteriosclerotic heart disease were examined by 64-slice spiral CT coronary artery imaging and coronary artery angiography during the 1-month hospitalization in the Department of Cardiology, Xuanwu Hospital, Capital Medical University from June 2005 to March 2006.METHODS:
280 segments of 28 patients were quantitatively analyzed coronary artery stenosis by selective coronary artery angiography and multi-slice spiral CT imaging based on eye-measurement diameter method. MAIN OUTCOMEMEASURES:
True positive, true negative, false positive, false negative, sensitivity, specificity, accuracy, positive predictive value, and negative predictive value of coronary artery stenosis were measured by multi-slice spiral CT imaging.RESULTS:
All 28 patients were included in the final analysis. Multi-slice spiral CT imaging showed that the sensitivity, specificity, positive predictive value, and negative predictive value were 46.5%, 97.6%, 86.8%, and 84.3%, respectively. If excluding the effect of 31 coronary segments with severe calcification, the sensitivity, specificity, positive predictive value and negative predictive value were 66.7%, 98.6%, 90.3% and 93.6%, respectively.CONCLUSION:
Multi-slice spiral CT imaging is simple, reliable, noninvasive and safe; moreover, it has a good potential for diagnosing especially excluding coronary arteriosclerotic heart disease, but still some limits.