Your browser doesn't support javascript.
loading
Application of LabSoCS efficiency calibration method in rapid analysis at lab under emergency monitoring of the nuclear incidents / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 163-166, 2011.
Artículo en Chino | WPRIM | ID: wpr-412719
ABSTRACT
Objective To explore the effeetiviness of the method of LabSOCS(Laboratory sourceless calibration software)efficiency calibration in laboratory rapid analysis for emergency monitoring of nuclear incidents.Methods The detection efficiency of three kinds of environmental samples in emergency monitoring Wag calculated bY using the LabSOCS efficiency calibration method,and compared with the values that were obtained by way of radioactive source calibration method.Results The maximum relative deviation of the detection efficiency between the two methods was less than 15%,and the values with relative deviation less than 5%accounted for 70%.Conclusions The LabSOCS efficiency calibration method might take the place of radioactive source efficiency calibration method,and meet the requirement of rapid analysis in emergency monitoring of the nuclear incidents.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2011 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2011 Tipo del documento: Artículo