A Comparative Study of Radiographic Images on Normal Anatomical Structures / 대한구강악안면방사선학회지
Journal of Korean Academy of Oral and Maxillofacial Radiology
;
: 283-297, 1999.
Artículo
en Coreano
| WPRIM
| ID: wpr-41604
ABSTRACT
PURPOSE:
To compare radiographic images of Digora system and Ektaspeed Plus film obtained from normal adults. MATERIALS ANDMETHODS:
Storage phosphor plate(SPP) was placed in a film holder behind Ektaspeed Plus film package without lead foil. The effect of film on SPP was studied in a separate in vitro experiment. Forty-seven sets of images were prepared for the evaluaton. The regions of interest(ROI) for evaluation were designated at seven sites including normal anatomical structures. The image quality for each ROI was evaluated on enhanced and unenhanced storage phosphor(SP) images and Ektaspeed Plus film.RESULTS:
Two film-SPP configurations showed significantly different grey levels at each step of the aluminum step wedge(p<0.05). The contrasts were comparable. Enhanced SP images were significantly superior to unenhaned images and film in all anatomical sturctures(p<0.01). The differences between unenhanced SP images and film were significant(p<0.05) except root canal and cortical bone on alveolar crest. For anatomical items, there were statistically significant difference among five observers(p<0.05).CONCLUSIONS:
The image quality of enhanced SP images were superior to Ektaspeed Plus film, and Digora system is potentially applicable to clinical diagnosis.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Radiografía Dental Digital
/
Cavidad Pulpar
/
Diagnóstico
/
Aluminio
Tipo de estudio:
Estudio diagnóstico
Límite:
Adulto
/
Humanos
Idioma:
Coreano
Revista:
Journal of Korean Academy of Oral and Maxillofacial Radiology
Año:
1999