Evaluation of the Tosoh HLC-723 G8 HbA1c Autoanalyzer / 임상검사와정도관리
Journal of Laboratory Medicine and Quality Assurance
; : 273-277, 2008.
Article
en Ko
| WPRIM
| ID: wpr-42693
Biblioteca responsable:
WPRO
ABSTRACT
BACKGROUND: We evaluated overall performance and analysis time of newly developed HLC-723 G8 (Tosoh corp., Tokyo, Japan) HbA1c autoanalyzerwhich utilizes cation-exchange HPLC method. METHODS: Linearity, precision, correlation with Variant II Turbo (Bio-Rad Laboratories, Hercules, CA, USA), analysis time, labile HbA1c (L-A1c) separation ability and stability of refrigerated sample at 4degrees C were evaluated. RESULTS: HLC-723 G8 showed good linearity between HbA1c 4.9-12.0% range (R2=0.9995). Within-run and total imprecision (CVs) were less than 1.0%. Correlation with Variant II Turbo and L-A1c separation ability were excellent. Analysis time per sample took 1.0 min. Stability of refrigerated sample at 4degrees C for at least 21 days was good. CONCLUSIONS: HLC-723 G8 showed good analytical performance and its rapid analysis time enables us to support outpatient diagnosis of diabetic patients efficiently.