Application of Monte Carlo software in estimating patients' radiation dose during CCTA and confirmation by anthropomorphic phantom study / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 152-154, 2014.
Artículo en Chino | WPRIM | ID: wpr-444857
ABSTRACT
Objective To evaluate the accuracy and efficiency of the Monte Carlo software in measuring the radiation dose to the patients who received the CCTA (Coronary Computed Tomography Angiography) examination.Methods A anthropomorphic chest phantom underwent CCTA using three scan parameters (tube voltage 80 kV,100 kV and 120 kV).Computer Software ImpactDose 2.0 was used to compute the chest organ dose on the basis of the three groups tube voltage CT scan characteristic,and the stimulation results of ImpactDose 2.0 software was verified by use of anthropomorphic phantom thermoluminescence dosimeter experiment method.Results For all the measured organs except for lung,the absorbed organ dose and effective dose of three groups of tube voltages of CCTA measured by the InpactDose 2.0 was lower than those as measured by anthropomorphic phantom study.The relative error of both methods was within ± 50%.Conclusions Monte Carlo software can be used to estimate the levels of radiation dose during CCTA examination with a tolerable error within the acceptable range.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2014 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2014 Tipo del documento: Artículo