Your browser doesn't support javascript.
loading
Development of a time-resolved fluorescence immunoassay for rapid determinat-ion of levels of cardiac troponin Ⅰantigen in human serum / 中国免疫学杂志
Chinese Journal of Immunology ; (12): 1093-1097, 2014.
Artículo en Chino | WPRIM | ID: wpr-454850
ABSTRACT
To establish a method to detect cardiac troponin I by using time-resolved fluoroimmunoassay ( TRFIA) and apply to the clinic.

Methods:

The assay were measured by TRIFA and double antibody sandwich method .Standard protocols were evaluated with the standard curve , the limit of detection , stability, precision and cross reaction .Healthy reference populations and clinical serum specimens were measured to established the reference interval and evaluated the perspective of the clinical application .

Results:

The standard curve was Y=7485 .878+1400.924 X with a correlation coefficient of 0.999.The limit of detection was 0.052 ng/ml.The intra-and inter-assay coefficients of variation ( CV) were all 0.05 ).There was significant difference in statistics compared before and after treatment with AMI ( P<0.001 ) .

Conclusion:

The TRFIA method for detecting cTnI achieves clinical application standards and may be used for the diagnosis and serosurveillance of acute myocardial infarction patients.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: Chinese Journal of Immunology Año: 2014 Tipo del documento: Artículo