Comparative analysis on results of different types of blood cell analyzer in blood stations / 国际检验医学杂志
International Journal of Laboratory Medicine
; (12): 747-748, 2015.
Article
en Zh
| WPRIM
| ID: wpr-460660
Biblioteca responsable:
WPRO
ABSTRACT
Objective To investigate the consistency of the results detected by different types of blood cell analyzer in the blood center .Methods On the basis of the calibration in each analyzer ,with the analyzer obtaining the excellent result in participating the external laboratory quality assessment hosted by the Ministry of Health as the reference ,the fresh blood samples were adopted to analyze other analyzers .Results For the analyzers after conducting the comparison ,the consistency and accuracy of the detection results were ensured .Conclusion After calibration in different types of blood cell analyzer ,the differences exist in the detection re‐sults .Periodically conducting the comparison among different types of instrument has very practical significance .
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
International Journal of Laboratory Medicine
Año:
2015
Tipo del documento:
Article