Your browser doesn't support javascript.
loading
Performance evaluation of Sysmex XN-9000 automated hematology analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 1373-1375, 2015.
Artículo en Chino | WPRIM | ID: wpr-463311
ABSTRACT
Objective To evaluate the performance of Sysmex XN‐9000 automated hematology analyzer .Methods According to international and domestic standards ,performance of analyzer was evaluated .Results The within‐batch and between‐batch preci‐sion ,carryover pollution rate ,linearity range and the accuracy of Sysmex XN‐9000 analyzer were all conform to related require‐ments .Leukocyte classification results compared with manual classification ,the correlation of neutrophil ,lymphocyte ,monocyte and eosinophil were fine ,but correlation of basophil was not very ideal .Conclusion The performance of Sysmex XN‐9000 analyzer could be satisfying ,could meet the needs of clinical inspection and diagnosis and treatment .

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: International Journal of Laboratory Medicine Año: 2015 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Guía de Práctica Clínica Idioma: Chino Revista: International Journal of Laboratory Medicine Año: 2015 Tipo del documento: Artículo