The study on the field environment suitability of multi-parameter patient monitor / 中国医学装备
China Medical Equipment
;
(12): 12-14, 2015.
Artículo
en Chino
| WPRIM
| ID: wpr-465463
ABSTRACT
Objective:
To study the environment suitability of multi-parameter patient monitor.Methods:
Choosing six different types of multi-parameter patient monitor from different manufactures, we tested six main parameters of the multi-parameter patient monitor when they had gone through the environmental tests, and the test limits come from field environment.Results:
Each of the samples was not up to the standards.Conclusion:
At present a clear most of the multi-parameter patient monitor in the market are designed and produced for using indoors, which are not suitable for using in field environment. We need to enhance the protection of the multi-parameter patient monitor against field environment, and improve the environment suitability.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
China Medical Equipment
Año:
2015
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS