Influence of dose rate reference control level on the design of accelerator shielding / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 139-141, 2015.
Artículo en Chino | WPRIM | ID: wpr-466251
ABSTRACT
Objective To explore the influence of dose rate reference control level on the design of accelerator shielding.Methods According to the standards of GBZ 126-2011,GBZ/T 201.1-2007 and GBZ/T 201.2-2011,two different shielding calculation methods,based on the dose rate reference control level and week dose control level,were used to calculate and compare the thickness of accelerator room shielding.Results Under the same condition of maximum weekly workload,the obtained results were different when the calculated dose rate reference control value was larger than dose rate reference control level (2.5 μSv/h).The maximum difference of shielding thickness reached 64 cm.Meanwhile,considering dose rate reference control level,the different accelerator rates could lead to different radiation shielding thickness.Conclusions The dose rate at reference point must be first calculated before calculating shielding thickness.The calculation should be made on the premise that dose rate reference control level is met.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Medicine and Protection Año: 2015 Tipo del documento: Artículo