Your browser doesn't support javascript.
loading
Long-term follow-up of right ventricular outflow tract septal pacing / 老年心脏病学杂志(英文版)
Journal of Geriatric Cardiology ; (12): 71-74, 2009.
Artículo en Chino | WPRIM | ID: wpr-472429
ABSTRACT
Objective Right ventricular outflow tract septum has become widely used us an electrode placement site. However, data concerning lead performances and complications for lead repositioning with this technique were scant. The purpose of this study was to observe long-term lead performances and complications of right ventricular outflow tract septal pacing and provide evidences for choosing an optimal electrode implantation site. Methods Thirty-six patients with septal active electrode implantation and 39 with apical passive electrode implantation were enrolled in this study. Pacing threshold, R-wave sensing, lead impedance, pacing QRS width and pacing-related compli-cations for two groups at implantation and follow-up were compared. Results There were higher pacing threshold and shorter pacing QRS width at implantation in the septal group compared with the apical group. There were no differences between the septal and the apical groups in pacing threshold, R-wave sensitivity, lead impedance and pace-related complication during a follow-up. Conclusions Right ventricular outflow tract septum could be used as a first choice for implantation site because it had long-term stable lead performances and no serious complications compared with the traditional apical site.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Journal of Geriatric Cardiology Año: 2009 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Journal of Geriatric Cardiology Año: 2009 Tipo del documento: Artículo