Your browser doesn't support javascript.
loading
Microstructure and function of Pix'L and other blunt-tip mixrocannulas for filler injection / 中华医学美学美容杂志
Article en Zh | WPRIM | ID: wpr-473035
Biblioteca responsable: WPRO
ABSTRACT
Objective To discuss the microstructure and properties of vascular puncture with a blunt needle Pix'L filling injection.Methods Scanning electron microscopy (SEM) and photoelectron spectroscopy were employed for the analysis of the different morphology and elements,and a rat model was established to study thrusts of different microcannulas and hypodermic needle when stabing vessels.Ultimately we analyzed the correlation between the morphology and puncture force of different microcannulas.Results The SEM images of microcannulas from different manufacturers were different,and revealed the different technics of different manufacturers.Puncture forces of different microcannulas corresponded to the cross-section and smooth degree of the tips.Conclusions The SEM morphology and puncture force of microcannulas from different manufacturers are different,which indicate the different characters and usability.Plastic surgeons should select to use them more judiciously.
Palabras clave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Medical Aesthetics and Cosmetology Año: 2015 Tipo del documento: Article