ABSTRACT
Objective To establish a new method for the rapid and sensitive determination of biapenem. Methods A simple, rapid and sensitive flow injection chemiluminescence(CL) method for the determination of biapenem was developed based on the fact that biapenem could enhance the CL of the luminol (LMN)-potassium ferricyanide reaction system under an alkaline condition. Results Under the optimal experimental conditions, we obtained a linear relationship between the CL intensity and the concentration of biapenem in the range of 0.02~0.6 mg/L with a detection limit of 7.3×10-3 mg/L. The relative standard deviation was 1.76% for the eleven-time determination of 0.2 mg/L biapenem. Conclusion The method shows high sensitivity, low detection limit and good repeatability. It has been successfully applied to the determination of biapenem for injection and the results are satisfactory.
Palabras clave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Journal of International Pharmaceutical Research Año: 2015 Tipo del documento: Article