Study of Stainless Steel Solid Phase Microextraction Fiber for Selective Absorption of Impurities in Gutter Oil / 分析化学
Chinese Journal of Analytical Chemistry
;
(12): 1422-1428, 2015.
Artículo
en Chino
| WPRIM
| ID: wpr-481837
ABSTRACT
A special solid phase micro extraction ( SPME ) fibre was successfully attained by coating with a special adsorptive material on a porous stainless steel support. And the porous stainless steel support was prepared by electrochemical corrosion for the first time. The optimized erosion condition in this research was as follows the current intensity of 30 mA, the stainless steel was etched in 2 mol/L hydrochloride solution for 1 min, then sonicated for 3 min; these procedures repeated 5 times. Meanwhile, the treated fiber coated 35 times in the sorbent solution with concentration of 5% ( m/V) in dichloromethane. The SPME fibre was applied to the analysis of gutter oil samples, and the performance of the SPME fibre for the headspace solid phase micro extraction of the exogenous impurities in gutter oil was specific, durable, stable and low-cost. The four detected trace level characteristic impurities in certain gutter oil were given as follows42 . 7 mg/kg for acetic acid, 21. 6 mg/kg 3-butenenitrile, 71. 8 mg/kg carbon disulfide and 2. 8 mg/kg allylisothiocyanate.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
Chinese Journal of Analytical Chemistry
Año:
2015
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS