ABSTRACT
Objective:To characterize the volatile compounds in 10 batches of disposable infusion sets and 6 batches of nasal can-nulas by GC-MS and determine the main odor-active compounds. Methods:The volatile components were extracted using a headspace sampler. An HP-5MS capillary column (30 m × 0. 25 mm,0. 25 μm) was adopted, and the qualitative analysis was performed by total ion chromatography ( TIC) of full scan with temperature programmer. Results:A total of 19 major volatile compounds were identified, which were hydrocarbon, alcohol and carbonyl compounds (such as aldehyde and ester). Based on the combination of odor test and GC-MS, the concentration of alcohol compounds (2-ethyl hexanol, 2-EH) had the most notable effect on the odor of samples. Conclu-sion:The samples with unacceptable order contain 2-EH with relatively high content, which should be paid more attention.