Study on the reliability of CardioChek PA for measuring lipid profile / 北京大学学报(医学版)
Journal of Peking University(Health Sciences)
;
(6): 523-528, 2016.
Artículo
en Chino
| WPRIM
| ID: wpr-493781
ABSTRACT
Objective:
To evaluate the clinical correlation between the CardioChek PA analyzer (CCPA)and a clinical laboratory reference method to use for screening program purposes.Methods:
Fasting blood samples were collected on 325 patients (age23 -86 years).One venous sample was col-lected using a serum tube for the evaluation on a Beckman reference analyzer.A second venous sample was collected in a lithium heparin tube and was evaluated on the CCPA analyzer.Linear regression analy-ses and Bland-Altman method were performed for each measured analytetotal cholesterol (TC),high density lipoprotein-cholesterol (HDL-C),triglycerides (TG)and low density lipoprotein-cholesterol (LDL-C).Results:
Our results demonstrated a good clinical agreement for TC,HDL-C,TG and LDL-C (97.0%,92.9%,92.4% and 83.7%)in comparison with the CCPA to the reference analyzer.The correlation coefficients were 0.875,0.81 3,0.91 0,0.864,respectively.P values all <0.001 .There was no significant difference in the detection rate of hyperlipidemia in TC,HDL-C and LDL-C.Conclu-sionWe have identified the pre-analytic phase as an important step to guarantee the quality of results and indicated that the CCPA is a reliable lipid point-of-care testing system that can be used for the appli-cation of clinical screening anywhere.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
Journal of Peking University(Health Sciences)
Año:
2016
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS