Your browser doesn't support javascript.
loading
Based on the teeth in vitro skull specimen to investigate CARE kV CT scan technology of decreasing radiation dose / 实用医学杂志
The Journal of Practical Medicine ; (24): 1813-1816, 2016.
Artículo en Chino | WPRIM | ID: wpr-494525
ABSTRACT
Objective Based on the teeth in vitro skull specimen to investigate the feasibility and accuracy of CARE kV CT scan for reducing radiation dose. Methods Of eight fixed and preserved skull teeth were scanned repeatedly in groups of A, B, C. In expose factors of 100 kVp/150 mAs in conventional scanning of group A, open CARE Dose 4D scan of group B and CARE kV scan of group C. We assessed the radiation dose in the group A, B and C, the average CT values of the images, noise, SNR, CNR and subjective ratings. Results Three groups of actual scan tube current, tube voltage and radiation dose were statistically significant (P 0.05). Conclusion CARE kV technology based on the use specimens tooth can reduce the radiation dose, protect image quality, and be consistent to the reports of previous researches.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: The Journal of Practical Medicine Año: 2016 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: The Journal of Practical Medicine Año: 2016 Tipo del documento: Artículo