Determination of Serum Concentration of Metoprolol by HPLC-Fluorescence Detection Method / 中国药房
China Pharmacy
;
(12)2007.
Artículo
en Chino
| WPRIM
| ID: wpr-531261
ABSTRACT
OBJECTIVE:
To establish a HPLC-fluorescence detection method for the determiniation of the serum concentration of metoprolol.METHODS:
The determination was performed on Symmetry C18 column,and the mobile phase consisted of acetonitrile-water-triethylamine -phosphoric acid(110∶390∶2.5∶1.6) at a flow rate of 0.8 mL?min-1.The excitation wavelength was 265 nm and the emission wavelength 298 nm,and the sample size was 20 ?L.RESULTS:
Good linearity was obtained for metoprolol over the range of 2.0~100.0 ?g?L-1 with correlation coefficient r=0.999 6.At low,medium and high concentrations,the average recoveries of metoprolol were 99.73%,98.21% and 99.38%,respectively.The intra-day RSD were 2.89%,2.36% and 1.32%,respectively,and the inter-day RSD were 3.73%,3.03% and 2.25% respectively.The lowest detectable limit was 1.0 ?g?L-1.CONCLUSION:
This method is precise,accurate,specific,simple yet with high recovery,and it is applicable for clinical monitoring of blood concentration and pharmacokinetic study.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Tipo de estudio:
Estudio diagnóstico
Idioma:
Chino
Revista:
China Pharmacy
Año:
2007
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS