Your browser doesn't support javascript.
loading
Enhancement of Image Contrast in Linacgram through Image Processing / 대한방사선종양학회지
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 345-354, 2000.
Artículo en Coreano | WPRIM | ID: wpr-54489
ABSTRACT

PURPOSE:

Conventional radiation therapy portal images gives low contrast images. The purpose of this study was to enhance image contrast of a linacgram by developing a low-cost image processing method. MATERIALS AND

METHODS:

Chest linacgram was obtained by irradiating humanoid phantom and scanned using Diagnostic-Pro scanner for image processing. Several types of scan method were used in scanning. These include optical density scan, histogram equalized scan, linear histogram based scan, linear histogram independent scan, linear optical density scan, logarithmic scan, and power square root scan. The histogram distribution of the scanned images were plotted and the ranges of the gray scale were compared among various scan types. The scanned images were then transformed to the gray window by pallette fitting method and the contrast of the reprocessed portal images were evaluated for image improvement. Portal images of patients were also taken at various anatomic sites and the images were processed by Gray Scale Expansion (GSE) method. The patient images were analyzed to examine the feasibility of using the GSE technique in clinic.

RESULTS:

The histogram distribution showed that minimum and maximum gray scale ranges of 3192 and 21940 were obtained when the image was scanned using logarithmic method and square root method, respectively. Out of 256 gray scale, only 7 to 30% of the steps were used. After expanding the gray scale to full range, contrast of the portal images were improved. Experiment performed with patient image showed that improved identification of organs were achieved by GSE in portal images of knee joint, head and neck, lung, and pelvis.

CONCLUSION:

Phantom study demonstrated that the GSE technique improved image contrast of a linacgram. This indicates that the decrease in image quality resulting from the dual exposure, could be improved by expanding the gray scale. As a result, the improved technique will make it possible to compare the digitally reconstructed radiographs (DRR) and simulation image for evaluating the patient positioning error.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Pelvis / Tórax / Posicionamiento del Paciente / Cabeza / Articulación de la Rodilla / Pulmón / Cuello Límite: Humanos Idioma: Coreano Revista: The Journal of the Korean Society for Therapeutic Radiology and Oncology Año: 2000 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Pelvis / Tórax / Posicionamiento del Paciente / Cabeza / Articulación de la Rodilla / Pulmón / Cuello Límite: Humanos Idioma: Coreano Revista: The Journal of the Korean Society for Therapeutic Radiology and Oncology Año: 2000 Tipo del documento: Artículo