Your browser doesn't support javascript.
loading
Matrix Modifiers Optimization for Determination of Chromium in Cosmetics by Graphite Furnace Atomic Absorption Spectrometry / 环境与健康杂志
Journal of Environment and Health ; (12)1993.
Artículo en Chino | WPRIM | ID: wpr-546090
ABSTRACT
Objective To select a matrix modifier,which can effectively eliminate the matrix interference in the determination of chromium in cosmetics by graphite furnace atomic absorption spectrometry(GFAAS).Methods The orthogonal experiment design was used to define the best operation parameter of graphite furnace.Vitamin C,(NH4)H2PO4,(NH4)2HPO4,NH4Cl and Mg(NO3)2 were added respectively as the matrix modifiers.The effect of five matrix modifiers was compared by precision test.Results When(NH4)H2PO4 was taken as the chemical modifier,the determination result was satisfied.The characteristic mass was 5.68?10-12 g,relative standard deviation were from 2.11% to 4.46%.The rate of recovery of two cosmetics samples were 95.2% and 102.01%.Conclusion(NH4)H2PO4 can increase the ashing temperature,decrease the atomization temperature,eliminate the matrix interference and the result has good precision and accuracy.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Journal of Environment and Health Año: 1993 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Journal of Environment and Health Año: 1993 Tipo del documento: Artículo