Determination of Fe,Cu,Zn,Mn in Pollen by Flame Atomic Absorption Spectrophotometry / 环境与健康杂志
Journal of Environment and Health
;
(12)1989.
Artículo
en Chino
| WPRIM
| ID: wpr-547650
ABSTRACT
Objective To establish a method for the determination of Fe, Cu, Zn, Mn in the pollen by flame atomic absorption spectrometry(FAAS).Methods Wet digestion-flame atomic absorption spectrophotometry was used to determine the content of Fe, Cu, Zn, Mn in the pollen.Results In the selected experimental conditions, the recovery rates of these elements were 98.5%-103.0% and the coefficient of variation was 1.13%-2.36%.The detection limits were 0.40 ?g/L for Fe, 0.08 ?g/L for Cu, 0.20 ?g/L for Zn, 0.14 ?g/L for Mn.The linear range was 0.00-2.00 mg/L.Conclusion FAAS is simple, fast, accurate for the determination of Fe, Cu, Zn, Mn in the pollen.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
Journal of Environment and Health
Año:
1989
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS