Purpose:

To analyze factors which influence the precision & accuracy of target positions in Alderson Head phantom and(or) patients.Materials and

Methods:

A target position simulator was used to determine the precision and accuracy of target localization while Alderson head phantom used to determine the overall precision and accuracy through the treatment procedure.

Results:

The overall precision and accuracy through the treatment procedure was found to be 1.72?0.60mm, and its contribution from CT localizing step, which was 1.4?0.3mm。

Conclusion:

Since there are more factors related to patients' treatment, the precision and accuracy of simulating in Alderson phantom is the best estimate for patient treatment.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiation Oncology Año: 1995 Tipo del documento: Artículo