Perfunnance evaluation on a latex enhanced immunoturbidimetric?_2-microglobulin kit / 中国基层医药
Chinese Journal of Primary Medicine and Pharmacy ; (12)2005.
Artículo en Chino | WPRIM | ID: wpr-560438
ABSTRACT
Objective To evaluate the performance of a latex enhanced immunoturbidimetric?2-microglobulin kit in clinical use. Methods The lower limit of detection,precision,accuracy and linearity range was determined on the automatic chemistry analyzer Hitachi 7170S. Results The lower limit of detection was 0. 09mg/L. The within-run precision was less than 4% while the between-run precision was less than 8%. The accuracy was 99.29 % and the linearity range was 0.2~18mg/L. Conclusion The?2-microgiobulin kit shows good performance in clinical use.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Primary Medicine and Pharmacy Año: 2005 Tipo del documento: Artículo