Your browser doesn't support javascript.
loading
Research on the influence of functional magnetic resonance imaging parameter to the quality of image in brain area / 中国医学装备
China Medical Equipment ; (12): 29-30,31, 2014.
Artículo en Chino | WPRIM | ID: wpr-572141
ABSTRACT

Objective:

To discuss the effect of the choice of brain functional MRI parameter on image quality.

Methods:

To select nine young volunteers for scanning brain functional MRI from February to December, 2012, and analyze the effect caused by parameters to the image quality.

Results:

The lower of the b value, the more level 1 image, the less level 3 image;the more of the number of diffusion sensitive gradient direction, the more level 1 image, the less level 3 image. The largest impact on image quality is the interlayer spacing, the number of diffusion sensitive gradient direction and the b value.

Conclusion:

The factors influencing the quality of brain functional MRI images including the interlayer spacing, the number of diffusion sensitive gradient direction and b value. In the parameter selection process, one should start from diagnostic purposes and weigh the pros and cons with overall consideration.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: China Medical Equipment Año: 2014 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: China Medical Equipment Año: 2014 Tipo del documento: Artículo