An entirely automatic system of nucleic acid amplification testing in blood screening and follow up study of positive donors / 中华检验医学杂志
Article en Zh | WPRIM | ID: wpr-585324
Biblioteca responsable: WPRO
ABSTRACT
Objective To establish the entirely automatic method of nucleic acid amplification testing (NAT) for blood screening and to study the feasibility of NAT.Methods Using entirely automated extraction method to extract nucleic acid , amplified and detected by Roche COBAS AMPLICOR system,evaluating the sensitivity and efficacy.Results The 95% limits of HBV DNA, HCV RNA/HIV-1 RNA tests by automation system were 38.9,16.4IU/ml and 20.4 copies/ml,95% Confidence Intervals were [21,323], [10.5,342] and [12,300] respectively.8 of 16 512 donations were PCR for HBV DNA positive,the DNA positive rate was 0.048%.7/8 donations were Anti-HBc positive,The last one was also converted positive.No positive HCV RNA and HIV RNA was detected. 3/6 following up samples seroconverted.Conclusions The entirely automatic system can be applied in blood screening.
Palabras clave
Texto completo: 1 Índice: WPRIM Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Screening_studies Idioma: Zh Revista: Chinese Journal of Laboratory Medicine Año: 2003 Tipo del documento: Article