The initial stability of CF/PEEK femoral stem after total hip replacement / 中华创伤骨科杂志
Chinese Journal of Orthopaedic Trauma ; (12)2004.
Artículo en Chino | WPRIM | ID: wpr-585384
ABSTRACT
Objective To explore the initial stability of a carbon fiber reinforced polyetheretherketone composite (CF/PEEK) stem in human femurs after total hip replacement. Methods 8 pairs of human cadaveric femurs were used to test the CF/PEEK prosthesis versus a conventional cementless stem of cobalt-chrome molybdemum alloy(CoCrMo). The bone-femoral prosthesis interface motion was measured in vitro using displacement transducers. Combined axial and torsional loads and simulated single-leg stance loading were applied. Subsidence, rotation and interface motion were measured with load cycles up to 1 000 N. Results The extent of the micro-motion in the proximal femoral region was larger than that in the distal femoral region. The CF/PEEK composite prosthesis showed less dynamic micro-motion in the proximal region compared with the CoCrMo alloy prosthesis stem, with the average difference being 42%. The extent of the distal micro-motion in the 2 groups was small but the difference was significant (P

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Orthopaedic Trauma Año: 2004 Tipo del documento: Artículo

Similares