OBJECTIVE To develop a new type of piezoelectric immunoglobulin microarray immunosensor for(determination) of immunoglobulin.METHODS The energy converters were 10MHz AT-cut quartz crystals with gold-coated electrodes.The monoclonal antibodies of immunoglobulin were immobilized onto the surfaces of(crystals) gold(electrodes) by SPA to study the best immobilization yield of the antibodies at different condition.Gold-(protein) A complex was highly stable and was believed to be Van der Waals in nature.We did experiments to find the best temperature-time condition about the binding of SPA and the gold surface and to explore the influence(about) the immobilization of antibody by the concentration and the pH of the antibody solution.RESULTS At the same antibody concentration, it had the highest immobilization yield of SPA at a low-temperature-long-time(combination).The most immobilization yield was obtained at the 5mg/ml concentration of antibody solution.The optimum pH for antibody immobilization was 7.2.CONCLUSIONS SPA for the immobilization of antibody is of(simple) operation,and time-saving.It is a good way to immobilize antibody on the gold electric surface of the quartz(crystal).