Your browser doesn't support javascript.
loading
Best Operation Mode of AP-960 Full-automated ELISA System / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)1993.
Artículo en Chino | WPRIM | ID: wpr-591619
ABSTRACT
Objective To compare the operation modes of AP-960 full automated ELISA system.Methods Two models were tested beginning with adding specimens(model 1) and beginning with adding enzymes(model 2).The following factors were compared the consumption of suckers,usage of time,result of test and the usual alarm during the test.Results The consumptions of suckers were 564 and 84 respectively.The needing time of the former mode was 125 minutes and that of the latter mode was 90 minutes.Conclusion The mode beginning with adding enzymes(model 2) is better for the current status of our clinical laboratory.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Medical Equipment Journal Año: 1993 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Medical Equipment Journal Año: 1993 Tipo del documento: Artículo