Study on error analysis during detecting aflatoxin B1 in food by using electrochemical immunosensor / 国际检验医学杂志
International Journal of Laboratory Medicine
;
(12): 947-948,951, 2014.
Artículo
en Chino
| WPRIM
| ID: wpr-598978
ABSTRACT
Objective To expand the application of electrochemical immunosensor during deleting aflatoxin B1 in foods and feeds through analyzing impacts of the time of antibody incubation and sample preparation .Methods T he double self-assembly immu-nosensor combined with aflatoxin B1 and carboxylated single-walled carbon nanotubes (SWNTs) was characterized by cyclic volta-mmetry and impacts of the time of antibody incubation and sample preparation methods were investigated .Results The signal in-creased gradually following the increasing time of antibody incubation and reached a plateau at 90 min and sample preparation meth-ods showed a comparatively large impact on results .Additionally ,the crude extractions purified through removing interfering com-pounds by immunoaffinity column could effectively eliminate the interference effects of sample matrix .Conclusion Deleting aflatox-in B1 by electrochemical immunosensor is characterized by various features ,such as fast ,simple and low detection limits .The pres-ent study shows that stability of the electrochemical immunosensor is affected by the time of antibody incubation and sample prepa-ration .
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
International Journal of Laboratory Medicine
Año:
2014
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS