Contents Determination of Rosin Acid in Rheumatoid Arthritis Tablet by HPLC-MS/MS / 中国药房
China Pharmacy
;
(12): 4299-4300,4301, 2015.
Artículo
en Chino
| WPRIM
| ID: wpr-605216
ABSTRACT
OBJECTIVE:
To establish a method for the content determination of rosin acid in Rheumatoid arthritis tablet. METH-ODSHPLC was performed on the column of ZORBAX SB-C18 with mobile phase of acetonitrile-0.1% formic acid(82∶18,V/V) at flow rate of 1.0 ml/min ,detection wavelength was 241 nm ,column temperature was 30 ℃ and volume injection was 10 μl. MS/MS column was ZORBAX SB-C18 with mobile phase of acetonitrile-0.1% formic acid(80∶20,V/V)at flow rate of 0.2 ml/min;column temperature was 30 ℃;volume injection was 0.5 μl. Ionization mode was ESI+,atomization gas pressure was 25 psi,gas flow as 8.0 L/min,capillary voltage was 4 000 V,capillary outlet voltage was 120 V,precursor ion was 303 m/z,scan range was 50-500 m/z and the gas temperature was 350 ℃.RESULTS:
The linear range of rosin acid was 2.5-100.0 μg/ml. RSDs of preci-sion,stability and reproducibility tests were lower than 2.0%,recoveries was 96.75%-98.11%(RSD=0.53%,n=6). CONCLU-SIONSThe method is simple,accurate and reproducible,and suitable for the content determination of rosin acid in Rheumatoid arthritis tablet.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Idioma:
Chino
Revista:
China Pharmacy
Año:
2015
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS