Your browser doesn't support javascript.
loading
Finite element analysis of the mechanical properties of the hip joint after hip surface replacement / 中国组织工程研究
Chinese Journal of Tissue Engineering Research ; (53): 4265-4270, 2017.
Artículo en Chino | WPRIM | ID: wpr-607730
ABSTRACT

BACKGROUND:

There are many researches on the finite element analysis of total hip arthroplasty, but the biomechanical distribution after hip surface replacement is little reported.

OBJECTIVE:

To analyze the biomechanical characteristics after hip surface replacement based on three- dimensional finite element analysis.

METHODS:

The three-dimensional finite element model after metal-to-metal hip surface replacement was established.The load of 350 N paralleling to the gravity line was loaded on the acetabulum. The stress distribution of the prosthesis and femur and the stress of the different regions of the femoral head were observed.RESULTS AND

CONCLUSION:

(1) The stress value of the acetabular cup was significantly higher than that of the acetabular body ((0.63± 0.34) vs. (1.89±0.67) MPa, P 0.05). (3) The stress value showed significant difference among femoral head regions (P <0.05), and the stress of the regions 6 and 8 was significantly higher than that of the other regions (P < 0.05). (4) These results show that after metal-to-metal hip surface replacement, femoral neck stress is concentrated, the stress of the femoral prosthesis is occluded, and the stress at the medial femoral neck as well as the junction of femoral neck prosthesis and bone is larger.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Tissue Engineering Research Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Journal of Tissue Engineering Research Año: 2017 Tipo del documento: Artículo