Comparison of two methods of tonometry in glaucoma patients:Goldmann applanation tonometer and non-contact tonometer / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science ; (12): 144-145, 2015.
Artículo en Chino | WPRIM | ID: wpr-636980
ABSTRACT
? AlM To compare the two methods for the measurement of glaucoma patients' intraocular pressure ( lOP) between Goldmann applanation tonometer ( GAT) and non-contact tonometer ( NCT) and find the laws of the two methods.?

METHODS:

The lOP of 108 glaucoma patients ( 206 eyes) were measured by GAT and NCT respectively.?

RESULTS:

ln 108 glaucoma patients, the average lOP of 206 eyes was 29. 77 ± 10. 27mmHg by GAT and 24. 59 ± 8. 58mmHg by NCT. There was significant difference between GAT and NCT (P<0. 01). The higher of lOP, the difference between GAT and NCT was greater.?CONCLUSlON The measurement results with NCT were lower than that of GAT. The higher of lOP, the difference between GAT and NCT was greater. lt's better to measure lOP by GAT for the glaucoma patients, in order to avoid the misdiagnosis and mistreatment of glaucoma.