Your browser doesn't support javascript.
loading
Quality Assurance of Intensity Modulated Radiation Therapy: Site-Specific Results of Eulji University Hospital / 대한방사선종양학회지
Article en Ko | WPRIM | ID: wpr-64853
Biblioteca responsable: WPRO
ABSTRACT
PURPOSE: To analyze our quality assurance (QA) data for intensity modulated radiation therapy (IMRT) according to treatment site and to possibly improve QA for IMRT in Hospital. MATERIALS AND METHODS: We performed QA on 50 patients (head and neck, 28 patients; Breast, 14 patients; Pelvis, 8 patients) for IMRT. The calculated dose from RTP was compared with the measured value film, gamma index, and ionization chamber for dose measurement in each case. RESULTS: The point dose measurement results in 45 of 50 patients showed good agreement with the calculation dose (+/-3%). The largest error measured thus far has been 3.60%, with a mean of only -0.17% (SD, 2.25%). Each treatment site showed an error rate of -0.13% (SD, 1.93%) for head and neck cases, -0.26% (SD, 2.79%) for breast cases, and -0.24% (SD, 2.44%) for pelvis cases. The gamma index verified with the error rate of head and neck cases (6%), breast (10%), and pelvis (6%), which corresponded to a tolerance of 3 mm (3% for the head and neck, 2%, for the breast 1% for the pelvis, and 0% in the region where the isodose curve was greater than 90%. CONCLUSION: We recognize the cause of errors for each treatment site of IMRT QA and so we maximize our efforts to reduce error and increase accuracy.
Asunto(s)
Palabras clave
Texto completo: 1 Índice: WPRIM Asunto principal: Pelvis / Mama / Cabeza / Cuello Límite: Humans Idioma: Ko Revista: The Journal of the Korean Society for Therapeutic Radiology and Oncology Año: 2011 Tipo del documento: Article
Texto completo: 1 Índice: WPRIM Asunto principal: Pelvis / Mama / Cabeza / Cuello Límite: Humans Idioma: Ko Revista: The Journal of the Korean Society for Therapeutic Radiology and Oncology Año: 2011 Tipo del documento: Article