OBJECTIVE:
The purpose of this study was to investigate the stability of mini-implants in relation to loading time.METHODS:
A total of 48 mini-implants (ORLUS, Ortholution, Korea) were placed into the buccal alveolar bone of the mandible in 8 male beagle dogs. Orthodontic force (200 - 250 gm) was applied immediately for the immediate loading group while force application was delayed for 3 weeks in the delayed loading group. For the subsequent loading periods (3, 6, 12 weeks), BIC (bone implant contact) and BV/TV (bone volume/total volume) and mobility test were carried out.RESULTS:
The immediate loading group showed no changes in BIC from 3 to 12 weeks, while the delayed loading group showed a significant increase in BIC between 3 and 12 weeks (p < 0.05). The BV/TV of the delayed loading group significantly increased from 6 to 12 weeks of loading (p < 0.05), while the BV/TV of the immediate loading group decreased from 3 to 12 weeks of loading. However, there was no significant difference in BV/TV between experimental groups. The mobility of the immediate loading group was not significantly different from that of the delayed loading group after 12 weeks of loading (p < 0.05).CONCLUSIONS:
These results showed that immediate loading does not have a negative effect on the stability of mini-implants compared to the early loading method in both the clinical and histomorphometric point of view.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Mandíbula
/
Compuestos de Mostaza Nitrogenada
Límite:
Animales
/
Humanos
/
Masculino
Idioma:
Coreano
Revista:
Korean Journal of Orthodontics
Año:
2008
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS