Your browser doesn't support javascript.
loading
Molecularly Imprinted Electrochemical Sensor for Determination of Cholesterol Based on Signal Amplification of Graphene / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 1535-1541, 2017.
Artículo en Chino | WPRIM | ID: wpr-658503
ABSTRACT
An electrochemical sensor was fabricated for determination of cholesterol ( ChO ) based on the molecularly imprinting and signal amplification of graphene. The sensor was prepared by electropolymerizing phenol on the surface of graphene modified glassy carbon electrode with ChO as template. The structure, properties and molecular imprinting effect of the imprinted membrane were studied by SEM, CV and DPV. The results showed that the sensor had good selectivity and high sensitivity to cholesterol. The calibration graph for the determination of cholesterol was linear in the range of 8. 0 × 10-8-2. 0 × 10-4 mol/L, with the detection limit (S/N=3) of 5. 6×10-8 mol/L. The sensor was applied to the determination of cholesterol in human serum samples with satisfactory results.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Analytical Chemistry Año: 2017 Tipo del documento: Artículo

Similares