Observation on the effect of the application of Diri urine test workstation and microscope in routine urine test / 中国生化药物杂志
Chinese Journal of Biochemical Pharmaceutics
;
(6): 425-426,429, 2017.
Artículo
en Chino
| WPRIM
| ID: wpr-659862
ABSTRACT
Objective To observe the effect of the application of Diri urine test workstation(consisting of FUS-200 urine analyzer and H-800 automatic urine analyzer) and microscope in routine urine test. Methods 378 cases of urine samples of patients in Longquan municipal people's hospital were selected randomLy. Urine routine test were carried out by three methods. The main test items were white blood cells (WBC) and red blood cell (RBC), the test results were compared. Results H-800 automatic urine analyzer the sensitivity of RBC was 94.3%, the specificity was 93.5%, the sensitivity of WBC was 85.4%, the specificity was 99.4%; FUS-200 urine analyzer the sensitivity of RBC was 85.7%, the specificity was 98.4%, the sensitivity of WBC was 92.7%, the specificity was 99.4%. Conclusion In order to get the most accurate results, we need to make a comprehensive analysis of the three kinds of test methods in order to provide a reliable basis for clinical diagnosis and treatment.